i-SENS, Inc.
This is a prospective, single arm, open-label, interventional, pilot study to evaluate the accuracy and precision of the continuous glucose monitoring (CGM) system of i-SENS, Inc., CareSens Air 3, in adult patients with type 1 diabetes.
Participants will wear two CareSens Air 3 and one Libre 3 Plus for 16 days without access to glucose values from the CareSens Air 3 devices while maintaining their original diabetes mellitus treatment. The participants will draw capillary blood at least 8 times a day using the SMBG system between study visits. Between insertion and removal of the CareSens Air 3 the participants will be in the clinic 4 times to obtain accurate glucose reading from the reference device under carefully controlled circumstances including induction of low and high blood glucose.
Inclusion Criteria: * Adults who are 18-65 years of age, inclusive * Patients with type 1 diabetes who use intensive insulin therapy (multiple daily injections (MDI) or continuous subcutaneous insulin infusion (CSII) including users of systems with continuous glucose monitoring (CGM) driven automatic adjustment of insulin doses for at least 3 months * Patients who voluntarily decide to participate in the study and provide written informed consent Exclusion Criteria: * The following abnormal skin or skin diseases or skin alterations at the CGM sensor attachment or insertion site: Severe psor…
Primary CareSens Air 3 device
Secondary CareSens Air 3 device