Philips Clinical & Medical Affairs Global
The main purpose of this study is to assess the equivalence between blood pressure measurements obtained using the Philips iNIBP system and those determined by the auscultatory reference method.
The main purpose of the study is to determine the equivalence in the mean difference (bias) and standard deviation (SD) between inflationary based non-invasive blood pressure (iNIBP) measurements (systolic blood pressure, diastolic blood pressure, mean arterial pressure) and reference auscultatory BP measurements, assessed using a same-arm sequential measurement method with dual independent observers.
Inclusion Criteria: * Participants must be 3 to 85 years of age. * Participants, parents or legal guardians of participants must be able to provide informed consent, parental permission, and/or assent form. * Participants that are between 7 and 17 (inclusive) years of age must provide assent to participate in the study. * Participants, parents or legal guardians of participants must be able to read or write in English and participants that are between the age of 7-17 must be able to understand English to provide assent. * Participant must be willing and able to comply with study procedures an…
The Philips inflationary based algorithm of the IntelliVue X3 Monitor will be used to measure systolic blood pressure, diastolic blood pressure, and mean arterial pressure compared to the ausculatory reference method.
Arctuva estimate
Actual pay is set by the research site.
Verify or correct the compensation shown for this listing.